M.C. Tomic, Z. Djinovic:
"Inline Liquid Concentration Measurement in Nanoliter Volume Using Fiber-Optic Low-Coherence Interferometry";
IEEE Sensors Journal, 8 (2008), 5; 587 - 592.



English abstract:
Abstract-A method for inline measurement of composition
of binary liquid mixture in a nanoliter volume is presented.
Low-coherence interferometric technique, based on fiber-optic
Mach-Zehnder interferometer, is applied for measuring the liquid
refraction index, from which its volume fractions are found. The
accuracy of volume fractions measurement, of about +/-0,5%, was
predominantly determined by the accuracy of reading the position
of mechanical scanner. The data rate of about 1.5 Hz was also
limited by mechanical scanning.

Keywords:
Concentration measurement, fiber-optic sensor, low-coherence interferometry, nanoliter volume


"Official" electronic version of the publication (accessed through its Digital Object Identifier - DOI)
http://dx.doi.org/10.1109/JSEN.2008.918700