B. Misini, W. Freinbichler, M.A Colivicchi, C. Ballini, K. Bisilimi, W. Linert, K. Tipton, L. Della Corte:
"Continuous monitoring of highly reactive oxygen radicals during in vivo microdialysis";
Journal of Neuroscience Methods, 251 (2015), S. 1 - 6.



Kurzfassung englisch:
Background: Terephthalate (TA2−), which reacts with highly reactive oxygen species (hROS) to form the fluorophor 2-hydroxy terephthalic acid (OH-TA) with a high selectivity, has been used for deter- mining hROS formation during in vivo microdialysis. Previously this involved collecting fractions of the microdialysate and determining the OH-TA formed after HPLC (the batch method). New method: This work reports the development and validation of a procedure for continuously deter- mining hROS formation during microdialysis. TA2− was added to the artificial cerebrospinal fluid (aCSF) perfusing medium to trap hROS. OH-TA formation was detected in real time with a sensitive fluores- cence detector equipped with a capillary flow cell that was coupled directly to the effluent stream of the microdialysis system.

Results: The behaviour of the system was assessed by comparison with the batch method and using a well-characterized animal model of excitotoxic damage, based on the application of high concentrations (1 mM and 500 􏰁M) of the non-NMDA glutamate receptor agonist kainate (KA) to the neostriatum. Data for the evoked release of taurine were also determined in these samples. No temporal difference between hROS and taurine release could be detected. Comparison with existing method(s): The flow method had a comparable sensitivity of hROS detection to the batch method. It was simpler, cheaper and less time-consuming than the batch method.

Conclusions: This direct system is convenient and technically undemanding. It should be useful for the rapid assessment of the hROS responses to neurotoxins and other compounds in microdialysis experi- ments in vivo

Schlagworte:
Highly reactive oxygen species (hROS) Terephthalate Fluorescence recording Capillary flow cell Online in vivo monitoring


"Offizielle" elektronische Version der Publikation (entsprechend ihrem Digital Object Identifier - DOI)
http://dx.doi.org/10.1016/j.jneumeth.2015.04.014