Confocal chromatic displacement sensors are versatile and precise sensors for measuring the distance to a single point. In order to obtain a 3D measurement device, this paper presents an integrated scanning sensor design that employs a tilting lens mechanism for manipulating the light path of the sensor. The optical implications of the design are analytically modeled and simulated. An experimental setup is constructed to evaluate the system design and to test its performance on a variety of samples. Results show good agreement with the simulations and modeling; with maximal tip/tilt angles of ±2.5°, the setup is capable of measuring a volume of 1.7x1.7x1mm3 with a lateral resolution of 24.8µm and an axial resolution of 3µm.


"Official" electronic version of the publication (accessed through its Digital Object Identifier - DOI)
http://dx.doi.org/10.1364/JOSAA.394212